Skip to main content

Table 1 Numerical values of fixed parameters [vascular resistance, zero-pressure filling volume (ZPFV), vascular compliance] used in the cardiovascular hemodynamic model (adopted from Heldt et al. [15] and Schreiner et al. [16])

From: Computational analysis of the effect of the type of LVAD flow on coronary perfusion and ventricular afterload

 

Resistance (PRU)

ZPFV (ml)

Compliance (ml/mmHg)

Reference

Right ventricle

0.003

50

1.2–20

[15]

Pulmonary arteries

0.08

90

4.3

[15]

Pulmonary veins

0.01

490

8.4

[15]

Left ventricle

0.006

50

0.4–10

[15]

Systemic arteries

 

715

2.0

[15]

Upper body 1

3.9

650

8

[15]

Upper body 2

0.23

  

[15]

Kidney 1

4.1

150

15

[15]

Kidney 2

0.3

  

[15]

Splanchnic 1

3.0

1,300

55

[15]

Splanchnic 2

0.18

  

[15]

Lower limb 1

3.6

350

19

[15]

Lower limb 2

0.3

  

[15]

AVC

0.01

250

25

[15]

IVC

0.015

75

2

[15]

SVC

0.06

10

15

[15]

LCX 1

27

0

0.2

[16]

LCX 2

2.7

  

[16]

LAD 1

27

0

0.2

[16]

LAD 2

2.7

  

[16]

CV

0.6

25

 

[16]

LCA

0.05

  

[16]

  1. AVC Abdominal vena cava, IVC inferior vena cava, SVC superior vena cava, LCX left circumflex artery, LAD left anterior descending artery, LCA left coronary artery, CV coronary vein